Button: Contact CNFButton: MultimediaButton: About CNF
Button: Getting StartedButton: PublicationButton: REU ProgramButton: Events & SeminarsButton: Education OutreachButton: TechnologiesButton: Lab Equipment

Button: Lab User


Suss MA6-BA6 Contact Aligner

SUSS MicroTech MA/BA 6 Contact Aligner

Suss MA/BA 6 Contact Aligner Backup: Garry Bordonaro , Michael Skvarla

Description:

This contact aligner uses 280-350 nm light to expose wafers up to 150 mm diameter. The system features an infrared camera for alignment to patterns on the backside of the wafer. The versatile mask holder allows both round and square plates as masks, and the sample plate accommodates small and odd-shaped substrates.


Alignment Mark Placement for Bottom Side Alignment (BSA)

* BSA objective have a range of travel 24 to 95mm apart.

* The 100mm and 150mm substrate chuck have two rectangles for objectives to view the substrate.

* 100mm Substrate Holder
* Each rectangle is 3cm by 2cm center symmetric.
* Rectangles have diagonals of endpoints (-4.5,1), (-1.5, -1) &
(4.5,1), (1.5,1)

* 150mm Substrate Holder
* Each rectangle is 4.5cm by 2cm, center symmetric
* Rectangles have diagonls of endpoints (-6.5,1), (-2,-1) & (6.5,1)

Capabilities:

  • Contact lithography system
  • Topside and Backside Alignment
  • Flood field exposure
  • Overlay accuracy to 1 um
  • Substrate size up to 150mm


Additional Resources:

Contact Lithography Alignment Keys
Contact Aligner Resist Matrix


Back to Top